Standards. There are insufficient data to support a technology standard for this topic.
Guidelines. There are insufficient data to support a technology guideline for this topic.
Options. In pediatric patients who require intracranial pressure (ICP) monitoring, a ventricular catheter or an external strain gauge transducer or catheter tip pressure transducer device is an accurate and reliable method of monitoring ICP.
A ventriculostomy catheter device also enables therapeutic cerebrospinal fluid (CSF) drainage.
Indications from Adult Guidelines. Recommendations from the adult guidelines (1) were not based on a level of evidence.
A ventricular catheter connected to an external strain gauge is the most accurate, low-cost, and reliable method of monitoring ICP. It also allows therapeutic CSF drainage. ICP transduction via fiberoptic or strain gauge devices placed in ventricular catheters provides similar benefits but at a higher cost.
Parenchymal ICP monitoring with fiberoptic or strain gauge catheter tip transduction is similar to ventricular ICP monitoring but has the potential for measurement drift.
Subarachnoid, subdural, epidural monitors (fluid coupled or pneumatic) and externally placed anterior fontanel monitors are less accurate.
The overall safety of ICP monitoring devices is excellent, with clinically significant complications (e.g., infection and hematoma) occurring infrequently.
In patients for whom ICP monitoring is indicated, a decision must be made as to what type of monitoring device to use. The optimal ICP monitoring device is one that is accurate, reliable, and costeffective and that causes minimal patient morbidity. We reviewed the scientific literature on ICP monitoring in children and adults and propose a ranking based on the currently available technology.
We searched Medline and Healthstar from 1966 to 2001 by using the search strategy for this question (see Appendix A) and supplemented the results with literature recommended by peers or identified from reference lists. Of 41 potentially relevant studies, two were used as evidence for this question (Table 1).
The scientific discussion of ICP monitoring technology is divided into the following pediatric and adult sections: A. ICP monitoring device accuracy and stability; B. optimal intracranial location of monitor; and C. complications.
Intracranial Pressure Monitoring Device Accuracy and Stability. There are no pediatric studies on this topic. In infants, external placement on an open anterior fontanel has been used, but there are no corroborative data on accuracy or stability of the device.
Optimal Intracranial Location of Monitor. There are no pediatric studies on this topic.
Complications. In a retrospective study of 49 pediatric patients with TBI between 2-16 yrs of age, Gambardella et al.
Most studies define infection as a positive CSF culture in ventricular and subarachnoid bolt monitors or a positive culture of the intracranial device. A better definition is bacterial colonization of the device rather than infection since there have been no reports in large prospective studies of clinically significant intracranial infections associated with ICP monitoring devices
Intracranial Pressure Monitoring Device Accuracy and Stability. The following information is quoted from the "Guidelines for the Management of [Adult] Severe Traumatic Brain Injury"
The Association for the Advancement of Medical Instrumentation has developed the American National Standard for Intracranial Pressure Monitoring Devices in association with a neurosurgery committee
According to the Association for the Advancement of Medical Instrumentation's standard, an ICP device should have the following specifications:
Pressure range: 1-100 mm Hg
Accuracy: ±2 mm Hg in range of 0-20 mm Hg
Maximum error: 10% in range of 20-100 mm Hg
Current ICP monitors allow pressure transduction by external strain gauge, catheter tip strain gauge, and catheter tip fiberoptic technology. External strain gauge transducers are coupled to the patient's intracranial space via fluid-filled lines, whereas catheter tip transducer technologies are placed intracranially. External strain gauge transducers are accurate and can be recalibrated, but obstruction of the fluid couple can cause inaccuracy. In addition, the external transducer must be consistently maintained at a fixed reference point relative to the patient's head to avoid measurement error.
Catheter tip strain gauge or fiberoptic devices are calibrated before intracranial insertion and cannot be recalibrated once inserted (without an associated ventricular catheter). Consequently, if the device measurement drifts and is not recalibrated, there is potential for an inaccurate measurement especially if the ICP monitor is used for several days.
There is potential for significant ICP measurement drift with fiberoptic pressure transduction and strain gauge pressure transduction in the parenchymal space. However, adult studies of catheter tip strain gauge ICP devices have demonstrated low or negligible drift over 5 days
Optimal Intracranial Location of Monitor. The following information is quoted from the "Guidelines for the Management of [Adult] Severe Traumatic Brain Injury"
A pressure transduction device for ICP monitoring can be placed in the epidural, subdural, subarachnoid, parenchymal, or ventricular location.
Historically, ventricular ICP is used as the reference standard in comparing the accuracy of ICP monitors in other intracranial compartments
The following statements ensue from review of the adult and pediatric literature:
Complications. The following paragraph is abstracted from the "Guidelines for the Management of [Adult] Severe Traumatic Brain Injury"
The complication rate for ICP monitoring is low. The most common complications are infection and loss of waveform. There are no pediatric reports documenting the incidence of significant brain injury, hemorrhage, or seizures as a result of ICP monitoring. There are no pediatric data on the use of prophylactic antibiotics to prevent infectious complications. In patients with ventriculostomy catheters who require continuous CSF drainage, ICP cannot be measured simultaneously. Although complications rarely produce long-term morbidity in patients, they can increase cost by requiring replacement of the monitor, and they can give inaccurate ICP readings. Each type of pressure transduction system and intracranial location of the monitor has a profile of potential complications. Calibration, monitoring for infection, and checking fluid coupled devices for obstruction are necessary tasks in maintaining an optimal ICP monitoring system.
In pediatric patients who require ICP monitoring, a ventricular catheter and/or an external strain gauge transducer or catheter tip pressure transducer device is an accurate and reliable method of monitoring ICP. A ventriculostomy catheter device also enables therapeutic CSF drainage. Clinically significant infections associated with ICP devices causing patient morbidity are rare and should not deter the decision to monitor ICP. The incidence of other complications, such as hemorrhage or seizures, is unknown, but the absence of reported incidents in the pediatric literature suggests that the incidence is probably low.
Parenchymal catheter tip pressure transducer devices measure ICP similar to ventricular ICP pressure but have the potential for measurement differences and drift due to the inability to recalibrate. These devices are advantageous when ventricular access is limited or unavailable or if there is obstruction in the fluid couple. There are no credible data (class III or better) on the accuracy of subarachnoid or subdural-coupled devices, epidural ICP devices, or externally placed anterior fontanel devices.
Prospective clinical studies in pediatric patients of the accuracy and complication rate of ventricular and intraparenchymal ICP measuring devices need to be performed. An industry or Food and Drug Administration supported national pediatric registry should be established to collect information on this and other issues in pediatric medicine.
The specification standard for pediatric ICP monitoring should include in vivo clinical ICP drift measurement. In vitro testing devices do not necessarily reflect clinical performance. Specifications for ICP devices should be reviewed in the context of what data are useful in the management of patients who require ICP monitoring.
A study of simultaneous parenchymal and ventricular ICP measurements using an accurate catheter tip transducer device in children would be useful. We must answer the question: Does parenchymal monitoring in or near a contusion site provide ICP data that improve intracranial pressure management and outcome compared with other sites (including contralateral sites) of ICP monitoring in children?
Recommendations for the use of prophylactic antibiotics, surgical techniques, ICP data collection, monitoring for complications, and timing for removal of ICP monitoring devices in children need to be developed. Further improvement in ICP monitoring technology should focus on developing an ICP device that can provide ventricular CSF drainage and parenchymal ICP measurement simultaneously. This would allow in situ recalibration and give accurate ICP measurements in case of fluid obstruction or when CSF is actively drained. Noninvasive measurements of ICP need to be developed.
Reference | Data Class | Description of Study | Conclusion |
---|---|---|---|
Gambardella et al., 1993 | III | Retrospective study of 49 patients that evaluated correlation between intraparenchymal Camino and ventriculostomy ICP catheters. | Good correlation between ICP measurements with the Camino vs. ventriculostomy catheter (r = .73-.89), with differences ranging from 1 to 4 mm Hg. |
Jensen et al., 1997 | III | Prospective uncontrolled case series of 98 patients with 12 ventriculostomy catheters and 37 intraparenchymal Camino catheters placed. | Infectious complication rate was 7% (all positive for Staphylococcus aureus). Loss of waveform occurred in 13% of catheters occurring at a mean of 9.5 days (range, 4-7 days). Data support low incidence of infection and mechanical failure. |